MD Series Scorpius
High-Performance Digital X-Ray and Fluoroscopy System for Clinical Imaging
High-precision radiography and fluoroscopy system designed for gastrointestinal and full-body diagnostic imaging. Equipped with a 3072×3072 flat panel detector, a powerful 84 kW generator, accurate positioning mechanisms, ergonomic interface, and advanced image processing tools for optimized diagnostics.
![]() | ![]() | ![]() |
---|---|---|
![]() | ![]() | ![]() |
The equipment is designed for radiographic and fluoroscopic examinations of the gastrointestinal tract (GIT) and for visual monitoring to obtain images for clinical diagnosis. It consists of an X-ray tube, diagnostic table, collimator, system console, flat panel detector, image acquisition workstation, high-voltage generator, and medical dry printer.
• Reference axis positioning in horizontal and vertical planes
• Visual display-based interface for intuitive human-machine interaction
• Programmable control technology ensures high precision, stability, and excellent anti-interference performance
• Frequency converter control provides smooth start/stop, stable and silent operation, and precise table positioning
• High-precision RVDT (Rotary Variable Differential Transformer) minimizes dynamic noise and extends mechanical life
• Tilt angle range: +90° ~ 0° ~ –25°, suitable for routine digestive organ exams
• Side switch design on the diagnostic table for more convenient local console positioning
• Dynamic flat panel detector enhances image clarity and contrast, simplifies lesion detection in radiography mode
• 3072×3072 matrix, up to 9.4 million pixels – ultra-high spatial resolution, fast response, excellent signal-to-noise ratio
• Large flat detector covers entire abdomen and chest without limitations
• Real-time image display with optimized software integration improves system efficiency
• Image analysis tools: angle, distance, shape-based ROI, brightness/contrast calibration, and stitching of saved images
• Amorphous silicon ensures stable performance, low environmental sensitivity, and resistance to crystallization at high temperatures
• Configurable technical exposure parameters
• Support for saving unfinished scans and images, image retransmission, reprocessing, reprinting, and patient data editing
• Patient type selection based on age and weight
Technical Specifications
• Max output power: 84 kW (84 kV, 1000 mA)
• Rated output: 80 kW (100 kV, 800 mA, 0.1 s)
• Max continuous output: 800 W (80 kV, 10 mA)
• Inverter frequency: up to 500 kHz
Voltage regulation (±10%):
• Radiography: 40–150 kV, 1 kV step
• Fluoroscopy: 40–125 kV, 1 kV step
Tube current regulation:
• Fluoroscopy: 0.5–10 mA (continuous), 5–40 mA (pulsed)
• Radiography: 10–1000 mA (±20%)
Anode input power:
• Large focus: 100 kW
• Small focus: 40 kW
• Heat capacity: 420 kJ (600 kHU)
• Rotation speed: 9700 rpm (180 Hz) / 2700 rpm (50 Hz)
• Heat dissipation: 1750 W (2465 HU/s)
Exposure timing:
• mAs: 0.1–1000 (±10% + 0.2 mAs)
• Time: 1–10,000 ms
• Focal spot: 1.2 mm (large), 0.6 mm (small)
Table movement and load:
• Tilt: +90° / –25° (±2°)
• Lateral movement: 220 ± 20 mm
• Elevation: 300 ± 20 mm
• Vertical lift: 800–1800 mm ± 50 mm
• Longitudinal detector travel: 1350 ± 50 mm
• Load capacity: 250 kg
• Footrest rotation: 360° (optional)
• Tabletop size: 220 × 72 cm
• Tube rotation: –90° to +180°
Detector specs:
• Pixel matrix: 3072 × 3072
• Pixel pitch: 139 μm
• Limiting resolution: ≥3.7 lp/mm
• Energy range: 40–150 kVp
• A/D conversion: 16-bit
• Frame rate: ≥30 fps (3×3 binning, 1024×1024, 42.7×42.7 cm FOV)
• LCD monitor: 23”
• Storage: 250 GB + 1 TB
• Operating system: Windows 10